Normal view MARC view

Intérêt de la mise en place de la recherche des gènes vanA et vanB par technique PCR en système clos (Xpert vanA/vanB Cepheid.) dans un laboratoire de microbiologie dans le cadre de la gestion d'une épidémie à Enterococcus faecium résistant aux glycopeptides (EfRG) [Article de Revue]

Implementation of vanA and vanB genes by PCR technique research interest in system (Xpert vanA/vanB CepheidR) closed in a laboratory of microbiology in managing an outbreak to Enterococcus faecium resistant glycopeptide (EfRG)
 BECLIN E ; DEKEYSER S ; Descamps, D
Pathologie-biologie. 2011/04; 59(2) : 73-78
Summary: OBJET : Face aux difficultés rencontrées lors de la gestion de deux épidémies à ERG, le laboratoire a étudié l'intérêt de rechercher les gènes vanA et vanB par PCR temps réel en système clos (Xpert vanA/vanB Cepheid.) pour améliorer le délai de réponse, la prise en charge des patients et le contrôle de la transmission croisée.METHODE : De mars à décembre 2009, 565 prélèvements ont été analysés par PCR, associé à un ensemencement systématique pendant deux mois de 75 échantillons puis culture uniquement des échantillons positifs en PCR pour les gènes vanA et/ou vanB.RESULTATS : La sensibilité et la valeur prédictive négative de la PCR étaient de 100 %. La spécificité a été évaluée en contexte et hors contexte épidémique : 69,3 et 76,8 % respectivement. Les variations interservices du taux de faux-positif en phase non épidémique sont moins importantes qu'en phase épidémique. Le taux global de faux-positif de 23,9 %.CONCLUSION : Cette technologie est d'intégration facile en routine : résultats en une heure pour quatre échantillons testés versus 72 heures pour la culture. Malgré son coût réactif, elle représente un outil diagnostique important à l'échelle d'un établissement : amélioration de la gestion des lits des zones de regroupement, transferts facilités, adaptation des mesures d'hygiène et de la prise en charge thérapeutique des patients. La culture reste cependant indispensable pour confirmer tout résultat positif obtenu en PCR et suivre épidémiologiquement la sensibilité des souches.(RESUME D'AUTEUR)

SUBJECT: The closed system PCR for the rapid detection of vanA and vanB genes (Xpert vanA/vanB Cepheid.) was evaluated in our laboratory, to improve the rapidity of the response and thus the management of patients and isolation measures during two GRE outbreaks.METHOD: From March to December 2009, 565 samples were analysed by PCR associated to bacterial culture initially for all samples for 2 months (n = 75), and thereafter for PCR-positive samples only.RESULTS: In this study, sensitivity and negative predictive values of the PCR were 100%. Specificity was evaluated in the presence and absence of outbreak: 69.3 and 76.8% respectively. The variability of false positive rates between units were lower in nonepidemic than during epidemic phase. The global false positive rate was 23.9%.CONCLUSION: This easy-to-use technology provides rapid results. four samples are tested in 1 h versus 72 h for culture. Despite its reagent cost, it represents an important hospital diagnostic tool: improvement of the management of cohorting areas and patient transfer between units, adaptation of isolation measures and treatments. However, culture remains necessary to confirm any positive result obtained by PCR and for epidemiological surveillance.(RESUME D'AUTEUR)
Publication
2011/04

Pages
73-78

Language
Français

Number of ref
8

Published in

ID notice
318793

Location Call number Status Date due Barcode
CCLIN Est
CClin Est
CHU de Nancy - Hôpitaux de Brabois
Rue du Morvan
54511 Vandoeuvre le Nancy Cedex
Tél : 03.83.15.34.73
cclin.est@chu-nancy.fr
http://www.cclin-est.org/
CCLIN Ouest
CClin Ouest
CHU Hôtel Dieu
CS 26419
35064 Rennes Cedex 2
Tél : 02.99.87.35.31
isabelle.girot@chu-rennes.fr
http://www.cclinouest.com/
CCLIN Paris Nord
CClin Paris-Nord
96 rue Didot
75014 Paris
Tél : 01.40.27.42.00  
karin.lebascle@sap.aphp.fr
http://www.cclinparisnord.org/
CCLIN Sud-Est
CClin Sud-Est
Hospices Civils de Lyon
Hôpital Henry Gabriel
20 route de vourles
69230 Saint-Genis-Laval
Tél : 04.78.86.49.50
cclinse@chu-lyon.fr
http://cclin-sudest.chu-lyon.fr/
CCLIN Sud-Ouest
CClin Sud-Ouest
CHU de Bordeaux
Hôpital Pellegrin - Bâtiment Le Tondu
33076 Bordeaux
Tél : 05.56.79.60.58
cclin.so@chu-bordeaux.fr
http://www.cclin-sudouest.com/

There are no comments for this item.

Log in to your account to post a comment.
Tags from this library:
No tags from this library for this title.
Log in to add tags.
    average rating: 0.0 (0 votes)
Languages: